• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Report text (for child record).Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported the gantry microphone was not working and the operator could not hear the patient at all times.If the operator is unable to hear the patient due to a failed component, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation.
 
Manufacturer Narrative
The customer discovered when attempting to adjust the volume of sound from the gantry, they noticed that no sound could be heard.The philips customer support engineer (cse) confirmed there was no report any patient impact or rescan as a result of this issue.The philips cse went to the site to investigate the reported problem that no sound could be heard coming from the gantry (patient).The cse evaluated the system and was able to confirm the customers allegation.The cse was able to determined that a failed breathing light assembly was the cause for the loss of communication from the gantry (patient) to the console (operator).The cse replaced the breathing light assembly to resolve the issue and the system was returned to full functionality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7557859
MDR Text Key109764957
Report Number1525965-2018-00408
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-