• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE Back to Search Results
Catalog Number 0035280
Device Problems Material Discolored (1170); Device Contamination with Chemical or Other Material (2944)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that black colored dust was found as the package was opened.It is unknown whether the dust was found inside or outside the device.
 
Manufacturer Narrative
The reported event was confirmed.Visual evaluation of the returned sample noted one opened (within unsealed original packaging), un-used bulb irrigation syringe present.No foreign material was noted within the packaging or the exterior of the syringe.A small (0.4mm^2) embedded black particle was observed on the bulb of the syringe.However, the size of the particle was deemed acceptable, as the specification stated loose or embedded foreign material shall not exceed aggregate total of 0.6mm^2.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "¿irrigation syringe - bulb type, 50cc.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Contains or presence of phthalates (dehp).Contains or presence of phthalates: di (2-ethylhexyl) phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks.Single use.Do not resterilize.Do not use if package is damaged.Latex-free.Manufacturer.Sterilized using ethylene oxide.Caution, consult accompanying documents.Authorized representative in the european community.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.¿.
 
Event Description
It was reported that black colored dust was found as the package was opened.It was unknown whether the dust was found inside or outside the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7566266
MDR Text Key110175703
Report Number1018233-2018-02024
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Catalogue Number0035280
Device Lot NumberNGBZ1875
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received06/04/2018
Supplement Dates Manufacturer Received08/26/2018
Supplement Dates FDA Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-