• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Air Leak (1008); No Apparent Adverse Event (3189); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although air may have been leaking from the red cpc connector on the driveline of the freedom driver, the freedom driver continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the patient stated he could feel "air leaking" at the red cpc connector.The hospital staff reported that they were unable to feel such a leak but did note they could hear air movement at the red cpc connector.There was no reported adverse patient impact.
 
Manufacturer Narrative
The customer-reported air leak in the red (left) cpc (colder products company) connector was not able to be reproduced or confirmed during investigation testing; therefore, a root cause for the reported air leak could not be conclusively determined.Visual inspection of the driver, including the drivelines and both cpc connectors, revealed no abnormalities, and the driver passed all incoming functional test requirements.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7566353
MDR Text Key110197196
Report Number3003761017-2018-00219
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/10/2018
Initial Date FDA Received06/04/2018
Supplement Dates Manufacturer Received05/10/2018
Supplement Dates FDA Received09/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
-
-