• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) CAPIO¿ SLIM; HOLDER, NEEDLE, GASTROENTEROLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) CAPIO¿ SLIM; HOLDER, NEEDLE, GASTROENTEROLOGIC Back to Search Results
Model Number M0068318261
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a capio¿ slim was used during a procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio carrier broke outside the patient.The procedure was completed with another capio¿ slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
A visual examination of the returned capio slim revealed that the carrier was broken.The broken section of the carrier was not returned by the customer.A functional examination was not performed due to the device condition (carrier broken).A review of the device history record (dhr) indicated that the device met all material, assembly, and product specifications at the time of release to distribution.However, the investigation concluded that the most probable cause for this event is design product enhancement since a modification would improve performance, function or appearance of the product.There is an investigation open to address the issue of "carrier broken.".
 
Event Description
It was reported to boston scientific corporation that a capio slim was used during a procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio carrier broke outside the patient.The procedure was completed with another capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIO¿ SLIM
Type of Device
HOLDER, NEEDLE, GASTROENTEROLOGIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7567982
MDR Text Key110034512
Report Number3005099803-2018-01721
Device Sequence Number1
Product Code FHQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/18/2021
Device Model NumberM0068318261
Device Catalogue Number831-826
Device Lot Number0021832211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/14/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-