ZIMMER BIOMET, INC. OXFORD TRIAL BEARING SM SZ 4/5; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - the event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the instrument fractured approximately one month prior and was given to the sales representative to return.It is unknown whether the trial fractured during a surgical procedure.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint sample was evaluated and the reported event was confirmed through physical evaluation.Visual inspection of the returned trial identified that the metal shaft had fractured off.The device history records were reviewed and no discrepancies relevant to the reported event were identified.The device had an approximate field life of eight (8) years and three (3) months with an unknown record of use.Based on the information available, the root cause is determined as normal wear during the instrument's field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD TRIAL BEARING SM SZ 4/5
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7568370
MDR Text Key110036099
Report Number0001825034-2018-03772
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
PP010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number32-420374
Device Lot Number115440
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-