• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. AVITENE; AGENT, ABSORBABLE HEMOSTATIC, COLLAGEN BASED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. AVITENE; AGENT, ABSORBABLE HEMOSTATIC, COLLAGEN BASED Back to Search Results
Catalog Number UNKAA010
Device Problem Insufficient Information (3190)
Patient Problems Abscess (1690); Foreign Body Reaction (1868); Inflammation (1932)
Event Date 09/17/2013
Event Type  Injury  
Manufacturer Narrative
Based on the information provided the root cause can be associated to the surgeon not removing all of the avitene flour following hemostasis causing the issue (inflamed area of tissue/abscess formation) that presented several months postoperative.As reported the mass was removed without further issue and the patient is reported to have been doing well following the revision procedure.Per the instructions-for-use "any excess avitene (mch) not removed at the time of surgery may either present itself as a (recurring) mass or a (space occupying) lesion or it may lead to a foreign body reaction that may present with or without clinical signs and symptoms as a recurring mass or lesion or postoperative abscess formation upon imaging." without a lot number a review of the manufacturing records is not possible.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Used on patient.
 
Event Description
It was reported that on (b)(6) 2013 the patient underwent a craniotomy to remove a glioblastoma (brain tumor).During the procedure, the surgeon used avitene for hemostasis.As reported, the surgeon waited several minutes before irrigating and aspirating the remainder of avitene flour.As reported, several months later on (b)(6) 2013 the patient had a follow up brain mri performed which reportedly showed a lesion/mass in the same space the brain tumor was removed.Preoperatively, the surgeon thought regrowth of the tumor.It is alleged that one week later on (b)(6) 2013 the patient underwent a revision procedure which included reopening of the skull incision that had been previously made and the surgeon reportedly discovered there was an inflamed area of tissue/abscess formation that had encapsulated left over avitene collagen.As reported, the tissue was removed and sent to pathology which was reportedly benign.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVITENE
Type of Device
AGENT, ABSORBABLE HEMOSTATIC, COLLAGEN BASED
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
MEDCHEM PRODUCTS, INC. -1223089
160 new boston street
woburn MA 01801
Manufacturer Contact
laura sundberg
100 crossings blvd.
warwick, RI 02886
4018258462
MDR Report Key7569158
MDR Text Key110072151
Report Number1213643-2018-01809
Device Sequence Number1
Product Code LMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P80002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKAA010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/15/2018
Initial Date FDA Received06/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-