Model Number MMT-397T
Device Problems Bent (1059); Electromagnetic Interference (1194); Insufficient Flow or Under Infusion (2182)
Patient Problem Vomiting (2144)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that she is receiving a blood glucose level of over 600 mg/dl.Customer treated the high blood glucose with a manual injection of insulin.Customer alleges having experienced elevated blood glucose for several days.Customer reports vomiting as a symptom of the high blood glucose.Customer alleges under delivery of insulin from her insulin pump.Customer reports exposing her insulin pump to an x-ray machine.Customer was able to troubleshoot during the call.Customer reported a bent infusion set cannula on her infusion set.The infusion set is expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM QUICK SET MMT-397T
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7570817
MDR Text Key110130230
Report Number2032227-2018-04470
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-397T
Device Catalogue NumberMMT-397T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age29 YR
Patient Weight84
-
-