Model Number DTBA2QQ |
Device Problem
Pacing Problem (1439)
|
Patient Problem
Complaint, Ill-Defined (2331)
|
Event Date 05/23/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient experiences an increase in heart rhythm each morning around 2:00 am.During trouble-shooting which device feature may be causing the issue, capture management was programmed off, but the issue remained unresolved so it is speculated either the p/r wave test or the atrial lead position check is the feature.Further trouble-shooting is planned and the device remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was later reported that the atrial lead position check was the suspected feature and the run-time for the feature was reprogrammed to waking hours to avoid disrupting sleeping hours.
|
|