• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  UNKNOWN HIP INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  UNKNOWN HIP INSTRUMENTS Back to Search Results
Catalog Number UNK HIP INSTRUMENTS
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a drill bit broke off into the bone behind the acetabular cup while drilling prior to placing a screw.The broken bit was inaccessible, without removing the cup.Surgeon opted to leave behind the drill bit in patient vs doing more harm to remove.The bit broke in use during surgery.The surgeon implanted a pinnacle porocoat sector ii shell.The 35mm flex drill bit broke as surgeon was prepping cup for a 6.5 x 35mm bone screw.Given the broken bit in the screw hole, the surgeon opted to place a screw in another hole and leave the broken bit in the patient.I spoke with the dr.And although unhappy about the incident, he felt the best course of action, was to leave it in place.He said the bit was fully surrounded by cancellous bone and showed little potential to migrate.He said removing the cup and attempting to dig the fragment of drill bit out, would have done more harm than good.He informed the patient of the event and will monitor it, going forward.No surgical delay.
 
Manufacturer Narrative
The instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP INSTRUMENTS
Type of Device
HIP INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7572907
MDR Text Key110269429
Report Number1818910-2018-61427
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP INSTRUMENTS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age61 YR
-
-