• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493926224300
Device Problems Kinked (1339); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/08/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr: a synergy ous mr 3.00 x 24mm stent delivery system (sds) was returned for analysis.An examination of the crimped stent revealed damage.Strut segments 6 and 7 from the distal end of the stent were damaged with struts lifted and pulled distally.The undamaged crimped stent outer diameter (od) was measured and is within the specification.The balloon cones were reviewed and there were no issues to note.The balloon wings were tightly wrapped and evenly folded.A visual and tactile examination of the hypotube revealed multiple kinks.An examination of the shaft polymer extrusion found no issues.There were no signs of damage or strain to the port bond.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
Reportable based on device analysis completed on 14-may-2018.It was reported that crossing difficulties and shaft kinked were encountered.The 100% stenosed target lesion was located in a mildly tortuous and moderately calcified coronary artery.A 3.00 x 24mm synergy¿ drug-eluting stent was advanced but failed to cross the lesion.When the device was removed, it was noted that the shaft was kinked.The procedure was completed with another synergy stent.No patient complications were reported and the patient's status was good.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7574759
MDR Text Key110276685
Report Number2134265-2018-04856
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/18/2019
Device Model NumberH7493926224300
Device Catalogue Number39262-2430
Device Lot Number0021565145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-