• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN G7 CUP; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN G7 CUP; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Detachment Of Device Component (1104)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: item # unknown unknown g7 liner lot# unknown, item # unknown unknown screw lot# unknown, unknown femoral head, unknown femoral stem.Reported event was confirmed by review of xrays which displayed liner asymmetrically seen medially.No evidence of abnormal radiolucency or loosening.The femoral head is laterally positioned within the acetabular shell possibly secondary to malfunction of the inner liner and osseous penetration of an acetabular fixation screw.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr's associated with this report: 0001825034 - 2018 - 03813 ; 0001825034 - 2018 - 03811.
 
Event Description
It was reported that patient underwent total hip arthroplasty; subsequently, the liner disassociated from the cup.No revision had taken place at time of reporting.Attempts were made to obtain further information; however, none was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN G7 CUP
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7574902
MDR Text Key110321618
Report Number0001825034-2018-03812
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PNI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-