MAUDE Adverse Event Report: COVIDIEN PROTEGE RX CAROTID STENT SYSTEM; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PROTEGE RX CAROTID STENT SYSTEM; STENT, CAROTID Back to Search Results
Catalog Number SECX-10-7-40-135
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the physician intended to use a protégé rx with a 6f guardwire temporary occlusion and aspiration system for treatment of a lesion in the patient¿s carotid artery.The ifu was followed and the device was prepped without issue.Pre-dilation was carried out.It was reported that when the physician began delivering the protégé stent, post delivery of the temporary occlusion and aspiration system the distal tip of the stent delivery system(sds) got caught on the gold marker of the guardwire.The stent failed to be delivered to the target location.The stent was removed without any issue and a non-mdt device was used to complete the procedure without further issue.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE RX CAROTID STENT SYSTEM
Type of Device
STENT, CAROTID
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7581478
MDR Text Key110453538
Report Number2183870-2018-00321
Device Sequence Number1
Product Code NIM
UDI-Device Identifier00821684038727
UDI-Public00821684038727
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P060001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/16/2019
Device Catalogue NumberSECX-10-7-40-135
Device Lot NumberA476863
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/10/2018
Date Device Manufactured06/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-