Model Number H7493926228270
Device Problems Bent (1059); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr.: synergy ous mr 2.75 x 28mm stent delivery system was returned for analysis.An examination of the crimped stent revealed damage.Strut segment 14 from the proximal end of the stent was lifted and pulled in a distal direction.The remainder of the stent was undamaged.The undamaged crimped stent od (outer diameter) was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and there were no issues to note.The balloon wings were tightly wrapped and evenly folded.A visual and tactile examination of the hypotube revealed multiple kinks.An examination of the shaft polymer extrusion found no issues.There were no signs of damage or strain to the port bond.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event Description
It was reported that stent damage occurred.A 2.75 x 28 synergy¿ drug-eluting stent was selected for use to treat a lesion.However, during preparation, it was noted that some stent struts were not well crimped on the stent delivery system.The device was never used inside the patient and procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7582170
MDR Text Key110473764
Report Number2134265-2018-04889
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/29/2019
Device Model NumberH7493926228270
Device Catalogue Number39262-2827
Device Lot Number0021718342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-