Model Number H7493911332270 |
Device Problem
Bent (1059)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/22/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device is a combination product.(b)(4).
|
|
Event Description
|
It was reported that stent damage occurred.The stenosed target lesion was located in the right coronary artery.A 2.75x32mm promus element ¿ drug-eluting stent was selected for use; however, it was noted that the stent struts were lifted.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: promus element,mr,ous 2.75x32mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified stent damage.Damage was noted to 3 central stent strut rows.The outer diameter of the undamaged section of the crimped stent was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found a hypotube kink 41mm distal to distal end of strain relief.A visual and tactile examination of the shaft polymer extrusion revealed no issues.A visual and microscopic examination of the tip identified tip damage.No other issues were identified during analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
|
Event Description
|
It was reported that stent damage occurred.The stenosed target lesion was located in the right coronary artery.A 2.75x32mm promus element drug-eluting stent was selected for use; however, it was noted that the stent struts were lifted.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
|
Search Alerts/Recalls
|