• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY PROMUS ELEMENT ¿; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY PROMUS ELEMENT ¿; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number H7493911332270
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event Description
It was reported that stent damage occurred.The stenosed target lesion was located in the right coronary artery.A 2.75x32mm promus element ¿ drug-eluting stent was selected for use; however, it was noted that the stent struts were lifted.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Manufacturer Narrative
Device evaluated by mfr.: promus element,mr,ous 2.75x32mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified stent damage.Damage was noted to 3 central stent strut rows.The outer diameter of the undamaged section of the crimped stent was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found a hypotube kink 41mm distal to distal end of strain relief.A visual and tactile examination of the shaft polymer extrusion revealed no issues.A visual and microscopic examination of the tip identified tip damage.No other issues were identified during analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that stent damage occurred.The stenosed target lesion was located in the right coronary artery.A 2.75x32mm promus element drug-eluting stent was selected for use; however, it was noted that the stent struts were lifted.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELEMENT ¿
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7584399
MDR Text Key110583715
Report Number2134265-2018-05366
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/06/2019
Device Model NumberH7493911332270
Device Catalogue Number39113-3227
Device Lot Number20910166
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-