• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493926232270
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that stent damage occurred.The 85% stenosed target lesion was located in the severely calcified mid to distal left circumflex artery.A 2.75 x 32mm synergy¿ drug-eluting stent was advanced but failed to cross the lesion due to calcification.It was noted that the stent structure was damaged.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
 
Event Description
It was reported that stent damage occurred.The 85% stenosed target lesion was located in the severely calcified mid to distal left circumflex artery.A 2.75 x 32mm synergy¿ drug-eluting stent was advanced but failed to cross the lesion due to calcification.It was noted that the stent structure was damaged.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
 
Manufacturer Narrative
Device avail.For eval, returned to mfr.On, device returned to mfr., device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.  device evaluated by mfr.: synergy ous mr 2.75 x 32mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified proximal stent damage.Damage was noted to the 3 most proximal stent strut rows.The undamaged crimped stent outer diameter was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found a hypotube kink 480 mm distal to the distal end of the strain relief.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and microscopic examination found damage to the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Manufacturer Narrative
(b)(6).(b)(4).
 
Event Description
It was reported that stent damage occurred.The 85% stenosed target lesion was located in the severely calcified mid to distal left circumflex artery.A 2.75 x 32mm synergy¿ drug-eluting stent was advanced but failed to cross the lesion due to calcification.It was noted that the stent structure was damaged.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
 
Manufacturer Narrative
Device evaluated by mfr.: synergy ous mr 2.75 x 32mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified proximal stent damage.Damage was noted to the 3 most proximal stent strut rows.The undamaged crimped stent outer diameter was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found a hypotube kink 480 mm distal to the distal end of the strain relief.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and microscopic examination found damage to the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that stent damage occurred.The 85% stenosed target lesion was located in the severely calcified mid to distal left circumflex artery.A 2.75 x 32mm synergy¿ drug-eluting stent was advanced but failed to cross the lesion due to calcification.It was noted that the stent structure was damaged.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7587261
MDR Text Key110629027
Report Number2134265-2018-05288
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2019
Device Model NumberH7493926232270
Device Catalogue Number39262-3227
Device Lot Number0021259877
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Initial Date Manufacturer Received 05/23/2018
Initial Date FDA Received06/11/2018
Supplement Dates Manufacturer Received05/23/2018
07/03/2018
Supplement Dates FDA Received06/11/2018
07/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age60 YR
-
-