• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SUNNEX, LLC SUNNEX MRI CELESTIAL STAR MOBILE SURGICAL LIGHT; DEVICE MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SUNNEX, LLC SUNNEX MRI CELESTIAL STAR MOBILE SURGICAL LIGHT; DEVICE MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number CS2050M
Device Problem Device-Device Incompatibility (2919)
Patient Problem Injury (2348)
Event Date 05/26/2018
Event Type  No Answer Provided  
Event Description
A power supply unit for a rolling surgical light was wheeled into zone 4 on a cart which was brought in too close to the magnet.The power supply unit was attracted to the bore of the 1.5t siemens aera, causing minor injury to a surgical technologist.The mri technologist had just completed the weekly acr phantom test and began to wheel out the mr exam table when the dbs team began to introduce the procedure equipment ito zone 4 (mr exam room).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUNNEX MRI CELESTIAL STAR MOBILE SURGICAL LIGHT
Type of Device
DEVICE MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
SUNNEX, LLC
charlotte NC
MDR Report Key7596154
MDR Text Key111090280
Report NumberMW5077806
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCS2050M
Device Catalogue NumberCS2050M
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/12/2018
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age66 YR
-
-