Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Information (3190)
|
Event Date 05/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the bearing trial fractured while the surgeon was trialing during the procedure.No other instruments were used to complete the procedure as trialing was complete.There were no other complications or delays.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).This report is being submitted to report additional information.Reported event was confirmed by review of actual device received.The device history record (dhr) was reviewed with the following deviations identified: one piece was scrapped at job step 0000.The product has a possible field life of over 13 years and shows heavy signs of wear, therefore root cause was determined to be wear and tear over a successful field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|