Model Number MZ5309 |
Device Problems
Fluid/Blood Leak (1250); Kinked (1339); Split (2537)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
|
|
Event Description
|
The customer reported that the extension tubing is forming an indentation/kink when clamped then splitting in the middle of the loop where the indentation was and leaking.The radiology department was injecting isovue contrast for a ct angiogram with a psi no greater than 325 and pushed the medication at a maximum of 6ml/sec when this happened.There was no patient harm.
|
|
Event Description
|
The customer reported that the extension tubing is forming an indentation/kink when clamped then splitting in the middle of the loop where the indentation was and leaking.The radiology department was injecting isovue contrast for a ct angiogram with a psi no greater than 325 and pushed the medication at a maximum of 6ml/sec when this happened.There was no patient harm.
|
|