• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Device Displays Incorrect Message (2591); No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a fault alarm while supporting a patient.The customer also reported that the patient was switched to a back-up freedom driver without any reported adverse patient impact.
 
Manufacturer Narrative
The driver in "as received" condition passed all test sections and pressure performance metrics associated with normotensive and hypertensive settings.Additionally, during an extended observation run test, the driver performed as intended and cardiac output remained within specifications.The customer-reported issue of a fault alarm could not be reproduced during investigation testing.The root cause of the customer-reported issue cannot be conclusively determined as the driver performed as intended with no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7603756
MDR Text Key111169955
Report Number3003761017-2018-00235
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/08/2018
Initial Date FDA Received06/14/2018
Supplement Dates Manufacturer Received06/08/2018
Supplement Dates FDA Received08/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age44 YR
-
-