During the procedure with the patient prepped and on the table, the stimulator did not pass the self test resulting in a clinically significant delay.At the beginning of the procedure, the ep-4 stimulator displayed ¿ep-4 is on¿ but did not pass the self test.Troubleshooting included power cycling the stimulator and checking all cable connections but the issue remained.A 60 minute delay occurred while the ep-4 stimulator was being replaced for the procedure to continue.The procedure was completed with no adverse consequences to the patient.
|
One workmate¿ claris¿ ep-4 cardiac stimulator was received for analysis.Visual inspection of the returned stimulator revealed no anomalies.The power on self-test and preliminary voltage measurements were performed which confirmed the returned stimulator passed the internal hardware self-test criteria.It was also confirmed that the stimulator communicated with the ep-4 touchscreen pc successfully upon self-test completion.The stimulator was then subjected to, and subsequently passed, an extended pacing and memory test & numerous power cycles without issue followed by a successful full functional test that met specifications.No self-test failures or communication faults were identified throughout testing as the returned stimulator functioned as intended during investigation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott and the investigation performed, root cause of the reported event could not be conclusively determined as the returned stimulator functioned as intended & no anomalies were identified throughout the investigation.
|