• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD PLASTIPAK¿ LUER-LOK¿ SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD PLASTIPAK¿ LUER-LOK¿ SYRINGE Back to Search Results
Catalog Number 309649
Device Problems Physical Resistance (2578); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when drawing up medication using bd plastipak¿ luer-lok¿ syringes difficult plunger movement was found, as stated when they "press the syringe plunger all the way down to make the syringe plunger slide more easily before use, but they still think it is very hard to work with." there was no report of exposure, injury or medical intervention.
 
Manufacturer Narrative
Due to an it issue beginning on (b)(6) 2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: device returned to manufacture: yes.
 
Event Description
It was reported when drawing up medication using bd plastipak luer-lok syringes difficult plunger movement was found, as stated when they "press the syringe plunger all the way down to make the syringe plunger slide more easily before use, but they still think it is very hard to work with." there was no report of exposure, injury or medical intervention.
 
Manufacturer Narrative
H.6.Investigation summary: two sealed packaged 3ml syringes confirmed to be from batch #7338532,(b)(4).And two sealed packaged 5ml syringes were received 7338674 (b)(4).They were visually evaluated.No visual defects were observed.The samples were tested for breakout and sustaining forces.All forces were well within the specification for the two products.The plunger movement was subjectively evaluated and was observed to be smooth as designed.No defects were confirmed in the samples received, therefore root cause is not determined.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Investigation conclusion: unable to determine a root cause.
 
Event Description
It was reported when drawing up medication using bd plastipak¿ luer-lok¿ syringes difficult plunger movement was found, as stated when they "press the syringe plunger all the way down to make the syringe plunger slide more easily before use, but they still think it is very hard to work with." there was no report of exposure, injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PLASTIPAK¿ LUER-LOK¿ SYRINGE
Type of Device
SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key7612331
MDR Text Key111853074
Report Number1213809-2018-00356
Device Sequence Number1
Product Code FMF
UDI-Device Identifier30382903096498
UDI-Public30382903096498
Combination Product (y/n)N
PMA/PMN Number
K980987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other
Type of Report Initial,Followup,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2022
Device Catalogue Number309649
Device Lot Number7338674
Initial Date Manufacturer Received 05/28/2018
Initial Date FDA Received06/18/2018
Supplement Dates Manufacturer Received05/28/2018
05/28/2018
Supplement Dates FDA Received08/02/2018
09/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-