Product event summary: the device and data files were returned and analyzed.Data files showed a system notice indicating that the refrigerant level was too low to continue (#50013) on the date of the case.Visual inspection of the catheter showed that the od of the balloon proximal bonding was within specification 0.155¿.The balloon was misshaped and the push button of the catheter was in the back position.The push button retracted during insertion has been identified as the cause of the insertion difficulty.As it caused the balloon material to fold up in itself increasing the od and there was crack on the folding sleeve.The insertion difficulty was reproduced.Smart chip verification indicated the catheter was not used.Without reprogramming the catheter, it was connected to the console and no system notices were received.The catheter failed the performance test due to a system notice indicating that the balloon is deflated (#22406) which was triggered in less than 2 minutes.Also, the inflation was not sustained for 2 minutes.A dissection and pressure testing showed a leak through the luer connector adhesive to the push button.Also, a guide wire lumen kink was inside the balloons from the tip of the catheter.In conclusion, the reported af/fc compatibility and achieve broken issues were not confirmed through data analysis but were confirmed during testing.The catheter failed the product return inspection test due to a balloon material folding up caused by retraction of the push button.Also, the broken push button luer.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryoablation procedure, the balloon catheter could not be inserted into the sheath.The balloon catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.The catheter was returned to the manufacturer, analyzed, and tested out of specification.
|