SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 52MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120152
Device Problems Metal Shedding Debris (1804); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Test Result (2695)
Event Date 08/30/2017
Event Type  Injury  
Event Description
It was reported a revision surgery of the left hip due to elevated cobalt and chromium levels and metallosis.
 
Manufacturer Narrative
It was reported a revision surgery of the left hip due to elevated cobalt and chromium levels and metallosis.As of today, device return and additional information has been requested for this complaint but has not become available.Since neither the underlying medical documents nor device part details were received for investigation no thorough medical investigation, manufacturing record review and assessment of the reported event can be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this patient's particular case, our investigation remains inconclusive.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the bhr head and cup were removed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.No further actions are required at this time.The available medical documents were reviewed.Although it was reported that the cobalt and chromium levels were elevated, neither the levels nor the lab reports were provided.The cystic mass and gray fluid and debris reported to be found intraoperatively are consistent with findings associated with metallosis; however without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported elevated cobalt and chromium levels, ¿metallosis,¿ and cystic mass containing gray fluid and debris found intraoperatively cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 52MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key7613182
MDR Text Key111445601
Report Number3005975929-2018-00209
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2015
Device Catalogue Number74120152
Device Lot Number10CW25958
Was Device Available for Evaluation? No
Date Manufacturer Received06/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74121146 FEMORAL HEAD, LOT 10CW25860.; FEMORAL HEAD COMPONENT/ UNKNOWN LOT NUMBER.; FEMORAL HEAD COMPONENT/ UNKNOWN LOT NUMBER
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age49 YR
-
-