SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 40/46; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122146
Device Problems Material Fragmentation (1261); Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Joint Dislocation (2374)
Event Date 04/28/2014
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed on right hip due to pain, dislocation, and metallosis.
 
Manufacturer Narrative
It was reported that a revision surgery was performed on right hip due to pain, dislocation, and metallosis.As of today, device return and additional information has been requested for this complaint but has not become available.Since neither the underlying medical documents nor device part details were received for investigation no thorough medical investigation, manufacturing record review and assessment of the reported event can be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this patient's particular case, our investigation remains inconclusive.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the hemi head, cup and sleeve were removed.A competitor stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the cup, hemi head and sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the hemi head and sleeve.Similar complaints have been identified for the cup and this failure will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.No further actions are required at this time.The available medical documents were reviewed.Although the intraoperative findings of stained fluid, corrosion, and stained hypertrophic synovitis are consistent with findings associated with metallosis, with the limited information provided, the root cause of the reported reactions cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 40/46
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key7613812
MDR Text Key111475574
Report Number3005975929-2018-00223
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup,Followup
Report Date 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2011
Device Catalogue Number74122146
Device Lot Number9063
Was Device Available for Evaluation? No
Date Manufacturer Received06/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74222300 MODULAR SLEEVE, LOT 09LW25119; BHR FEMORAL HEAD/ UNKNOWN LOT NUMBER.; HEMI HEAD 74122540, LOT 08HW18317; BHR FEMORAL HEAD/ UNKNOWN LOT NUMBER.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age68 YR
-
-