MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G146
Device Problems High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/10/2017
Event Type  malfunction  
Manufacturer Narrative
At this time, the device remains in service.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) exhibited right atrial (ra) noise, oversensing, inappropriate atrial tachy response (atr) episodes, impedance measurements of 1300 ohms, and episodes of pacemaker mediated tachycardia (pmt).The oversensing appeared to be from minute ventilation (mv) signals.The device was reprogrammed to turn off the mv feature.No adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7614560
MDR Text Key111515020
Report Number2124215-2018-06539
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534577
UDI-Public00802526534577
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/26/2017
Device Model NumberG146
Other Device ID NumberINOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0185; 4136; 4677; G146; T175
Patient Age64 YR
-
-