Model Number G148
Device Problems Electrical /Electronic Property Problem (1198); Over-Sensing (1438); Ambient Noise Problem (2877); Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/05/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and right atrial (ra) lead exhibited noise and intermittent oversensing resulting in inappropriate atrial tachy response (atr) mode switches.Review of stored device memory noted fluctuating ra pacing impedance measurements with the highest measurement reaching 1,200 ohms.Boston scientific technical services (ts) reviewed the episodes and discussed that the noise was consistent with oversensing related to the minute ventilation (mv) feature and recommended programming the respiratory rate trend (rrt) off.Rrt was programmed off.No adverse patient effects were reported.This product remains implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7614682
MDR Text Key111514421
Report Number2124215-2018-06376
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/10/2016
Device Model NumberG148
Other Device ID NumberINOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 04/04/2018
Initial Date FDA Received06/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0292; 4135; 4674; G148
Patient Age65 YR
-
-