MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G148
Device Problems High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/04/2018
Event Type  malfunction  
Manufacturer Narrative
At this time, the product remains in service.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that inappropriate atrial tachy response (atr) episodes were stored on this cardiac resynchronization therapy defibrillator (crt-d).Review of the events indicated noise consistent with minute ventilation sensor oversensing, on the atrial channel of this crt-d.The atrial lead was another manufacturer's lead.There were a couple of spikes in atrial pace impedance up to 800 ohms, and an alert was also recently triggered for a high out of range atrial impedance.The longest period of oversensing that was noted, was 17 seconds; however this patient is not pacemaker dependent and is only paced 3% of the time.The patient is scheduled to have the respiratory rate trend feature programmed off.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7614959
MDR Text Key111659893
Report Number2124215-2018-08826
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/02/2018
Device Model NumberG148
Other Device ID NumberINOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age41 YR
-
-