• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
The report of the autopulse platform (sn (b)(4)) displaying a "system error, out of service, revert to manual cpr" error message was confirmed during functional testing and archive data review.Root cause was determined to be a faulty processor board and drivetrain motor.The platform failed the initial functional testing due to system error (latch error 139 - unable to hold compression position) was observed upon powering up the device.No physical damage was observed during visual inspection.The autopulse platform is a reusable device and was manufactured in 2006 and is (b)(4) years old, well beyond the expected service life of five years.The archive data review indicated system error 139 (unable to hold compression position) around the reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
The autopulse platform (sn (b)(4)) displayed a system error, out of service, revert to manual cpr message during patient use.The user was unable to resolve the issue.Following this, the user immediately performed manual cpr on the patient.No known impact or patient consequence was reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7617406
MDR Text Key111996422
Report Number3010617000-2018-00640
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/21/2018
Initial Date FDA Received06/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-