The report of the autopulse platform (sn (b)(4)) displaying a "system error, out of service, revert to manual cpr" error message was confirmed during functional testing and archive data review.Root cause was determined to be a faulty processor board and drivetrain motor.The platform failed the initial functional testing due to system error (latch error 139 - unable to hold compression position) was observed upon powering up the device.No physical damage was observed during visual inspection.The autopulse platform is a reusable device and was manufactured in 2006 and is (b)(4) years old, well beyond the expected service life of five years.The archive data review indicated system error 139 (unable to hold compression position) around the reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
|
The autopulse platform (sn (b)(4)) displayed a system error, out of service, revert to manual cpr message during patient use.The user was unable to resolve the issue.Following this, the user immediately performed manual cpr on the patient.No known impact or patient consequence was reported.No additional information was provided.
|