• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS Back to Search Results
Catalog Number 762165
Device Problems Leak/Splash (1354); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2017
Event Type  malfunction  
Manufacturer Narrative
Bd received samples and photos from the customer facility for investigation.The samples and photos were evaluated and the customer's indicated failure mode with the incident lot was observed.Additionally, retention samples were selected for evaluation, and the customer's indicated failure mode was not observed.
 
Event Description
It was reported that paxgene® blood rna tubes were collapsed.No serious injury or medical intervention reported.No report of blood exposure to mucous membrane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7620551
MDR Text Key112062285
Report Number9617032-2018-01137
Device Sequence Number1
Product Code NTW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
DEN050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2018
Device Catalogue Number762165
Device Lot Number6194598
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-