Model Number 3100 |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/23/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Vyaire medical reference number (b)(4).At this time, vyaire medical has not received the suspect device from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the piston kept stopping while the device was in use on a patient.The circuit lost pressure each time causing the piston to stop.There was no patient harm/injury associated with the reported issue.The customer reported calibrating the driver displacement indicator (ddi) board and completing the performance check.The customer reported that the device is working to specifications and cancelled the requested onsite field service.
|
|
Manufacturer Narrative
|
Results of investigation: this unit was field serviced by a vyaire medical authorized field service technician.The service technician calibrated the ddi board and performed performance checks on the ventilator, and the ventilator is working within specifications.
|
|