• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number H7493925116270
Device Problems Bent (1059); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/26/2018
Event Type  Injury  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr.: the complaint device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
 
Event Description
It was reported that stent damage post deployment occurred.Vascular access was obtained via the femoral artery.The 75% stenosed, 15mm x 2.75mm, eccentric, de novo target lesion with a significant bend of > 45 and < 90 degrees was located in the moderately tortuous and mildly calcified left anterior descending artery.A 16 x 2.75mm promus premier¿ drug-eluting stent was implanted to treat the lesion.However, during post dilatation using a non-bsc balloon catheter, it was noticed that the proximal part of the stent got deformed.Another stent was then deployed and completed the procedure.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7621367
MDR Text Key111748085
Report Number2134265-2018-05427
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/03/2019
Device Model NumberH7493925116270
Device Catalogue Number39251-1627
Device Lot Number0020855081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient Weight51
-
-