• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number H7493952824220
Device Problems Bent (1059); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event Description
It was reported that stent damage occurred.During unpacking of a 2.25x24mm promus premier¿ drug-eluting stent, it was noted that a stent strut was pulled up.The device never entered the patient's body and the procedure was completed with another of the same device.No patient complications were reported.
 
Manufacturer Narrative
Device evaluated by mfr.: the stent delivery system was returned for analysis.A visual examination of the stent found that the first proximal stent row was damaged with stent struts lifted.The undamaged crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.A visual and microscopic examination of the tip found no damage.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the device found no issues along the mid shaft, inner or outer shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event Description
It was reported that stent damage occurred.During unpacking of a 2.25x24mm promus premier¿ drug-eluting stent, it was noted that a stent strut was pulled up.The device never entered the patient's body and the procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7623307
MDR Text Key111905869
Report Number2134265-2018-05510
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729844822
UDI-Public08714729844822
Combination Product (y/n)N
PMA/PMN Number
P110010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2019
Device Model NumberH7493952824220
Device Catalogue Number39528-2422
Device Lot Number21644719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/20/2018
Supplement Dates Manufacturer Received06/28/2018
Supplement Dates FDA Received07/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-