• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® URINE COLLECTION CUP; SPECIMEN TRANSPORT AND STORAGE CONTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® URINE COLLECTION CUP; SPECIMEN TRANSPORT AND STORAGE CONTAINER Back to Search Results
Catalog Number 364941
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Pma / 510(k) #: there is no 510(k) for this device as it is manufactured outside the us and not sold in the us.Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for missing needle with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a situation analysis.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for missing needle with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a situation analysis has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Rationale: based on an assessment of severity and frequency, it was determined that a situation analysis is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Event Description
It was reported that bd vacutainer® urine collection cups were missing needles.No serious injury or medical intervention reported.No report of blood exposure to mucous membrane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® URINE COLLECTION CUP
Type of Device
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7625204
MDR Text Key112063381
Report Number9617032-2018-02467
Device Sequence Number1
Product Code KDT
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2019
Device Catalogue Number364941
Device Lot Number8010514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/26/2018
Initial Date FDA Received06/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-