VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100A |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Information (3190)
|
Event Date 05/24/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).At this time, vyaire medical has not received the suspect device from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the despite which setting the blender knob was set to, the delivered fio2 was only 60%, while the device was in use on a patient.The customer attached another blender to the ventilator and the blender worked correctly.The patient was on extracorporeal membrane oxygenation and the customer did not provide any further information regarding the patient's condition in relation to the reported issue.
|
|
Manufacturer Narrative
|
Results of investigation: vyaire medical was not able to verify the customer's reported issue during testing and evaluation.The system was supplied with air and oxygen, and the unit pressurized as intended passing the circuit leak test.The unit was cycled for 1 hour with no performance issues, however, the blender was tested at a setting of 60%, and the measured fio2 reading was 55.6%.At all other settings, the fio2 was within specification.The reported issue of the blender staying at 60% regardless of the knob setting was not duplicated and the report of the loose knob was not duplicated.The reading of the fio2 reading at a setting of 60% was due to the front stem calibrating being off.Vyaire adjusted the front stem to correct the reported issue.
|
|
Search Alerts/Recalls
|
|
|