Model Number 801188 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) replaced the oxygen (o2) sensor.
|
|
Event Description
|
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) displayed a 'service gas system message' error message.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Event Description
|
Per clinical review: during a bypass procedure on (b)(6) 2018, the perfusionist received a 'service gas system' message on the messaging bar of the heart lung machine (hlm).The team just continued the procedure with the equipment and had no concern with the sweep gas flow nor the percentage of fraction of inspired oxygen (fio2) being delivered to the patient.After the procedure, the field service representative (fsr) did come in and replace the oxygen (o2) sensor.This perfusion group does leave their hlms plugged into the main power and powered on at all times.The incident did not delay the surgical procedure and there was no harm or blood loss associated with the occurrence.
|
|
Manufacturer Narrative
|
Please disregard previously reported information as this complaint has been determined to be not mdr reportable.The device met specification and performed as intended, therefore did not malfunction.
|
|