• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) replaced the oxygen (o2) sensor.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) displayed a 'service gas system message' error message.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Event Description
Per clinical review: during a bypass procedure on (b)(6) 2018, the perfusionist received a 'service gas system' message on the messaging bar of the heart lung machine (hlm).The team just continued the procedure with the equipment and had no concern with the sweep gas flow nor the percentage of fraction of inspired oxygen (fio2) being delivered to the patient.After the procedure, the field service representative (fsr) did come in and replace the oxygen (o2) sensor.This perfusion group does leave their hlms plugged into the main power and powered on at all times.The incident did not delay the surgical procedure and there was no harm or blood loss associated with the occurrence.
 
Manufacturer Narrative
Please disregard previously reported information as this complaint has been determined to be not mdr reportable.The device met specification and performed as intended, therefore did not malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7628859
MDR Text Key112003844
Report Number1828100-2018-00325
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)120905
Combination Product (y/n)N
PMA/PMN Number
K153376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/11/2018
Initial Date FDA Received06/22/2018
Supplement Dates Manufacturer Received07/12/2018
09/19/2018
Supplement Dates FDA Received08/06/2018
09/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-