Model Number G151
Device Problems High impedance (1291); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2018
Event Type  Injury  
Manufacturer Narrative
At this time, the product remains in service.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that during a normal device replacement, a code 1005 was recorded when the other manufacture's lead was hooked up to this new cardiac resynchronization therapy defibrillator (crt-d).This error is indicative of a high shocking lead impedance, greater than 145 ohms.The lead to header connection was checked and there were no issues.Fluoroscopy revealed no issues.A 5 joule commanded shock was delivered, to rule out interference, and was unable to shock, due to a code 1005.The other manufacturer's device did not note any impedance issues, prior to the removal of the lead.It is believed that the lead was fractured, during the device change.The lead was capped and replaced.The exact cause of the issue remains unknown.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7630914
MDR Text Key112083768
Report Number2124215-2018-11026
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534638
UDI-Public00802526534638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/24/2020
Device Model NumberG151
Other Device ID NumberDYNAGEN CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 05/08/2018
Initial Date FDA Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1888TC; G151; MISMATCH
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age41 YR
-
-