• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Break (1069); Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
Belt (b)(4) was returned and evaluated at the distributor.The reported problem (damaged connector) was confirmed.Upon evaluation of the electrode belt, the latches on the trunk cable connector were damaged, creating an insecure connection with the monitor.The root cause of the damaged connectors is excessive force.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor reported that a patient's belt was damaged and would not securely latch to the monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
gabrielle salazar
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key7631162
MDR Text Key112205706
Report Number3008642652-2018-05542
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/22/2018
Initial Date FDA Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-