GUIDANT CRM CLONMEL IRELAND COGNIS; IMPLANTABLE CHF PULSE GENERATOR Back to Search Results
Model Number N108
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2018
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Surgical intervention was performed and the crt-d was explanted and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This device was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Review of the device memory indicated that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COGNIS
Type of Device
IMPLANTABLE CHF PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7631296
MDR Text Key112093066
Report Number2124215-2018-11232
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/27/2012
Device Model NumberN108
Other Device ID NumberCOGNIS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/08/2018
Initial Date FDA Received06/22/2018
Supplement Dates Manufacturer Received07/03/2018
Supplement Dates FDA Received10/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
N108
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age73 YR
-
-