The device was not returned to olympus for evaluation.The root cause of the reported event could not be conclusively determined at this time; however, based on similar reported events the operator¿s technique during the use of the device could not be ruled out as a contributing factor to the reported event.The instruction manual provides warning and caution statements in an effort to mitigate the risk of patient harm and equipment damage which states, ¿examine this device prior to use.Do not use if damage is found.Always keep sheaths parallel to one another when assembling and disassembling.If the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip.A broken tip, or fragments of a damaged tip, can potentially pass through the outer sheath and into the patient.If drag or resistance is encountered during assembly or disassembly, stop/align working element and sheath parallel to one another before proceeding.¿.
|
Olympus was informed that during the middle of a transurethral resection of a prostate procedure, plastic pieces at the distal tip of the sheath broke off and fell into the patient¿s bladder.Three pieces were suctioned out of the patient with irrigation.The procedure was delayed by thirty minutes.The intended procedure was completed using the same device.There was no patient injury reported and there was no additional procedure/treatment was required.In addition, the device was inspected prior to procedure with no anomalies observed.The device is reprocessed utilizing steam sterilization (autoclave).
|
The device was returned to olympus for a physical evaluation.A visual inspection of the as received condition found the black colored ceramic insulation tip at the distal end of the device was broken and missing.A microscopic inspection revealed that the adhesive is still present inside the distal area where the insulation tip is attached.The remaining portion of the insulation tip inside the device has minor cracks and jagged edges.In addition, there are multiple scratches at the distal tip of the device, small deep indentations along the outer shaft tube and the red orientation dot on the shoulder screw is missing.However, the device passed the functional tests as the locking ring, release button and scope passage work appropriately.
|