WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990601
Device Problems Improper or Incorrect Procedure or Method (2017); Overcorrection (3006)
Patient Problem No Code Available (3191)
Event Date 04/21/2018
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device was out of production prior to the september 24, 2014 udi regulation date.(b)(4).
 
Event Description
A doctor reported incorrect programming of the cylinder axis during a lasik procedure.Residual astigmatism was treated.Upon follow up, patient was very satisfied with the result of the first eye and has already planned a retreatment of the second eye.The treatment rather than being programmed on the axis of 83 had been incorrectly entered with the axis of 173.The result was the astigmatism was increased by about 3 diopters more.Both the surgeon and the patient seem pretty quiet in managing this type of correction of user error.
 
Manufacturer Narrative
The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on company's acceptance criteria.The system history shows no abnormalities that could have contributed to this event.Contact with the clinical application specialist stated that the treatment was programmed on the axis 83 and has been incorrectly entered to the axis 173.The result was that the astigmatism was increased about 3 diopters more.The clinical application specialist confirmed it was an user error.The root cause for the reported event could be identified as use error in typing the incorrect axis.A product malfunction is excluded.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7633570
MDR Text Key112160071
Report Number3003288808-2018-01171
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/25/2018
Supplement Dates Manufacturer Received08/29/2018
Supplement Dates FDA Received09/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age20 YR
-
-