• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 09H39-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2018
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.An evaluation is in-process.
 
Event Description
The customer observed a falsely elevated platelet result while using the cell-dyn emerald analyzer.The customer provided the following results data (normal range is 140-440 k/ul) on (b)(6) 2018: sid (b)(6) : initial 1136, retest 335 10e3/ul.No adverse impact to patient management was reported.No additional patient information was made available.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, field service review, a search for similar complaints, and a review of labeling.The customer data indicated that the results generated flags which alert the operator to verify the results.Field service evaluated the analyzer and completed preventative maintenance.No additional issues have been documented since completion of maintenance.Tracking and trending did not identify an adverse trend.Labeling was reviewed and found to be adequate.Based on the available information no product deficiency of cell-dyn emerald analyzer, list number 09h39 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key7635007
MDR Text Key112754042
Report Number2919069-2018-00032
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740020088
UDI-Public00380740020088
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09H39-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2018
Initial Date FDA Received06/25/2018
Supplement Dates Manufacturer Received06/28/2018
Supplement Dates FDA Received07/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-