Model Number 97715
Device Problems Failure to Deliver Energy (1211); Device Operates Differently Than Expected (2913)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 01/02/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacture representative regarding a patient implanted with an implantable neurostimulator (ins) for failed back surgery syndrome.It was reported that the ins stim level changed from 9.6 ma to 0ma after the rep switched from their tablet to the patient's controller.It was discussed that this may have been due to the patient leaning forward with adaptive stim activated.It was also reported that the patient's therapy had been turning off by itself since implant, but the patient would be able to turn it back on again.It was reviewed that the patient take notes when this issue occurs.It was also reviewed that the ins was probably not turning off due to battery depletion as the patient is able to turn therapy back on afterwards.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was reported that the cause of the loss of therapy was not reported.The patient is going to take notes if this issue happens again.No further information was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7637375
MDR Text Key112300338
Report Number3004209178-2018-14290
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2018
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/26/2018
Date Device Manufactured12/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-