Device is a combination product.(b)(4).Device evaluated by mfr: stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent found damage to proximal stent rows 1-2 with stent struts lifted and pulled distally.The undamaged crimped stent outer diameter was measured and was within specification.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and do not appear to have been subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and the inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|