MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; CAT Back to Search Results
Model Number OPT312
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the returned complaint opt312 cannula was visually inspected.Results: visual inspection revealed that the left tube was stretched and damaged where it joins the patient interface.Although the tubing was stretched, the metal spring was still intact and attached to the interface: the glue bond was also still intact.Conclusion: the damage to the flexitube was most likely caused by an excessive pulling force.All optiflow junior cannulae are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are currently taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior nasal cannula.They also state the following: do not stretch or crush tube.Ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.- appropriate monitoring must be used at all times.
 
Event Description
A hospital in (b)(6) reported that the tubing of an opt312 optiflow junior nasal cannula was cracked and stretched after three days of use.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key7640370
MDR Text Key112850758
Report Number9611451-2018-00549
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT312
Device Catalogue NumberOPT312
Device Lot Number2100350983
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-