• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1711K
Device Problem Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that o ring was missing.Blood glucose was 6.7 mmol/l.Troubleshooting was performed.Ring of the reservoir was unable to lock in place when inserted into insulin pump.Advised the insulin pump will need to be replaced.Advised to discontinue use of the pump and revert to a back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with missing retainer ring and reservoir tube lip o-ring.Unable to perform displacement test and p-cap / reservoir will not lock properly due to missing retainer.
 
Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0958-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1711K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7640701
MDR Text Key112551181
Report Number3004209178-2018-85466
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169999763
UDI-Public(01)00643169999763
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1711K
Device Catalogue NumberMMT-1711K
Device Lot NumberHG2764S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/17/2018
Initial Date FDA Received06/27/2018
Supplement Dates Manufacturer Received09/14/2018
10/14/2019
Supplement Dates FDA Received10/03/2018
05/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-0958-2020
Patient Sequence Number1
-
-