• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problem Disconnection (1171)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/24/2018
Event Type  Injury  
Manufacturer Narrative
Manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 3.Reference mfr.Report#3006705815-2018-01496; reference mfr.Report# 1627487-2018-06041.It was reported the patient's ipg was unable to be set into mri mode.The patient underwent surgical intervention on (b)(6) 2018 wherein the leads were found to be out of the ipg header.The leads were reinserted in to the ipg header which resolved the issue.Intraoperative diagnostics revealed normal impedances on the leads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
taruna sharma
6901 preston road
plano, TX 75024
9725269635
MDR Report Key7642219
MDR Text Key112478131
Report Number3006705815-2018-01495
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/06/2020
Device Model Number3186
Device Lot NumberA000053176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/14/2018
Initial Date FDA Received06/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192 92), SCS ANCHOR
Patient Outcome(s) Other;
Patient Age42 YR
-
-