• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problems Syncope (1610); Ventricular Tachycardia (2132); Palpitations (2467)
Event Date 05/31/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that a syncopal patient was admitted to the hospital on (b)(6) 2018 due to ventricular tachycardia.The patient also experienced palpitations.During hospitalization, device interrogation conveyed loss of capture for the left ventricular lead.A chest x-ray was performed, which revealed left ventricular lead dislodgement.The ventricular tachycardia was unrelated to the dislodgement and loss of capture.Hospitalization was prolonged due to the dislodgement.Programming and medication dosage changes were made.The patient was discharged on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7651272
MDR Text Key112757267
Report Number2017865-2018-09497
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Model Number1458Q/86
Device Lot NumberA000046159
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age56 YR
-
-