• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.021
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the silver knob of the attachment device dropped off.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay in a planned surgical procedure, or if a spare device was available for use.There was no human patient involvement as this device is for veterinary use only.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
The date device returned to manufacturer was documented as 6/13/2018 on the initial report and has been updated as 6/14/2018.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device coupling on the tool side was not functioning and defective.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITTAL SAW ATTACHMENT
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7654368
MDR Text Key112854384
Report Number8030965-2018-54747
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819107182
UDI-Public(01)7611819107182(11)041123
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.021
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Initial Date Manufacturer Received 06/13/2018
Initial Date FDA Received07/02/2018
Supplement Dates Manufacturer Received07/03/2018
Supplement Dates FDA Received07/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-