• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE Back to Search Results
Catalog Number 0035280
Device Problems Contamination (1120); Device Contamination with Chemical or Other Material (2944)
Patient Problems No Patient Involvement (2645); No Information (3190)
Event Date 06/11/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the user found foreign material, which appeared to be trash, in the syringe.The device was not used.
 
Event Description
It was reported that the user found foreign material, which appeared to be trash, in the syringe.The device was not used.
 
Manufacturer Narrative
The reported event was confirmed as manufacturing related.Visual evaluation of the returned sample noted one opened, unused bulb syringe present.A small black particulate, resembling lint or detritus, measuring 0.1mm^2 was observed within the barrel of the syringe barrel.The presence of filth is considered out of specification, which states that no filth is allowed.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: ¿irrigation syringe - bulb type, 50cc this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Contains or presence of phthalates (dehp) contains or presence of phthalates: di(2-ethylhexyl)phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks single use do not resterilize.Do not use if package is damaged.Latex-free.Manufacturer.Sterilized using ethylene oxide.Caution, consult accompanying documents.Authorized representative in the european community.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7664457
MDR Text Key113454960
Report Number1018233-2018-02624
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0035280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Initial Date Manufacturer Received 06/11/2018
Initial Date FDA Received07/05/2018
Supplement Dates Manufacturer Received08/30/2018
Supplement Dates FDA Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-