• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX30012UX
Device Problems Mechanical Problem (1384); Device Dislodged or Dislocated (2923)
Patient Problems No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Event Date 06/11/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
One resolute onyx drug eluting stent was intended to be used to treat a non-tortuous and moderately calcified lesion with 80% stenosis in the mid rca.There was no damage noted to the device packaging.There were no issues removing the device from the hoop.The device was not inspected.Negative prep was performed with no issues.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during an attempted deployment.An angioplasty of the distal rca was performed prior to the stent dislodgement.The stent was not removed.No patient injury is reported.
 
Manufacturer Narrative
No difficulties noted when removing the protective sheath.The inflation device remain on neutral pressure during delivery of the device.An attempt was made to snare the dislodged stent but it was unsuccessful.The stent was lost in the patient anatomy.The angioplasty of the distal rca had no effect on the dislodgement of the stent.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7669082
MDR Text Key113631421
Report Number9612164-2018-01659
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169556973
UDI-Public00643169556973
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/16/2020
Device Catalogue NumberRONYX30012UX
Device Lot Number0008954928
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2018
Initial Date FDA Received07/09/2018
Supplement Dates Manufacturer Received07/16/2018
08/08/2018
10/02/2018
Supplement Dates FDA Received08/14/2018
10/03/2018
10/22/2018
Date Device Manufactured01/16/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient Weight97
-
-